The "Product FAQs" column collects products related questions which MGI customers inquiries frequently. And we organizes FAQs which will be released in the future to help you understand and use MGI products better.
Q1:What is the product positioning of MGISP-Smart 8?who is the target customer?
A brief introduction of MGISP-Smart 8:
1. Pipettes: 8 flexible pipetting channels;
2. Modules: temperature control module, shaker module, magnet module ,heater shaker module and PCR module etc.
3. Comsumables: tips, tubes, plates and reservoirs etc.
4. Software: intuitive graphical desk and process editor, support parallel processes and cherry picking
It is a flexible, scalable and user friendly automated liquid handler which could be used as a worksation to perform various applications in the fields of genomics, proteomics, drug discovery, cellomics and others.
Typical customers are pharmaceutical and biotech companies, as well as institutions constituting the worldwide bio-pharma market. it's intended for clinical and in-vitro diagnostics applications as well.
Q2:What's the difference between Smart 8 and SP-100/960?
The main differences between MGISP-Smart 8, MGISP-100 and MGISP-960 are pipettes, hardware and library preparation throughput. In addition, MGISP-100 and MGISP-960 are mainly sold in standard configuration, while Smart 8 can be sold in standard configuration or customized according to customer's application requirements and consumables.
Q3:What configurations/modules are available for the MGISP-Smart 8? How to identify what I need?
4 standard configurations: Config.1, Config.2, Config.3, Config.4,
2 hosts: Host with HEPA, Host without HEPA
12 carriers: Tube carrier A/B/C(Suitable for tubes)、Suspended Tip Carrier A(Suitable for tips)、Plate Carrier A/Combi. Carrier A(Suitable for microplates)、Single Well Reservoir Adapter A(Suitable for single well Reservoirs) and Combi. Carrier B/H/I/G/J(Contain functional modules)
5 optional function modules: Shaker Module A、Magnetic Module A/B、Robot Scanner A、Deck Scanner A
8 adapters: Tube adapter A/B、Single Well Reservoir Adapter A、PCR Plate Adapter A/B、Deepwell Plate Adapter A、Multi-segment Adapter A-C
Please contact local FAS or PM for indentify the configurations/modules in a certain scenarios.
Q4:Dose MGISP-Smart 8 have liquid level detection function?
MGISP-Smart 8 has Capacitive Liquid Level Detection (cLLD) . During the workflow, If the volume of the tube/well is not enough or the tube is empty, the system will prompt an validation error.
Q5:What are the SP-Smart 8 pipettes' accuracy and precision?
Pipetting range: 1μL-1000μL
Accuracy:1 μL: CV≤8%;200 μL: CV≤1%;1000μL: CV≤1%
Precision: 1μL≤±10%;200 μL≤±2%;1000μL≤±1%
Q6:Any MGI-SP Smart 8 parameters I could proposed for tender evaluation criteria?
Size: W: 1410mm,H:799mm,D:970mm (without HEPA module)
Net weight: Approximately 220 kg (485 lb, without carries)
Voltage: 200-240V,50/60Hz
Rated power: 1500VA ~3500VA(not including modules on operation deck and optional modules)
Working environment:
Ambient temperature:15℃-35℃
Relative humidity:20%RH-80%RH, No condensation
Air pressure range:80-106kPa
There are 4 types of consumables available, tips, tubes, microplates and reservoirs, as follows:
Tips:We recommend the use of MGI's T50, T200, T1000 conductive and non-conductive tips.
Tubes:0.5mL, 0.6mL, 1.5mL, 2mL, 5mL, 6mL, 10mL, 50mL centrifuge tubes or sample tubes.
Microplates:PCR Plates, Deepwell Plates and other 96-well SBS standard plates ,as well as Break-away PCR Plate, etc.
Reservoirs:50mL/100mL Single Well Reservoir made in MGI
For more information about the consumables, please refer to the brochure on MGI's website.
MGISP-Smart 8 provides various solutions to customer , it is able to accelerate workflows such as library preparation, nucleic acid extraction and reagents dispensing and preparation, cherry picking , sample pooling and normalization. We are continuing to work with our partners to expand our solution list,such as cell biology and synthetic biology.
For library preparation, three applications including WGS, WES and RNA library preparation can be carried out on Smart 8, as an open platform it can be compatible with MGI's kits and kits from third-party.
The Smart 8 minimizes aerosol and cross-contamination by using the HEPA module and filter tips.
1) The HEPA module continuously refreshes the air within the instrument, reducing aerosol contamination.
2) MGI's filter tips decrease the risk of aerosol contamination during the worklow.
It differs from MGISP-960 mainly in throughput and flexibility
(1)The Smart 8 support nucleic acid extraction for any number of samples/run from 8-96 , and it is more flexible and automated, allowing directly pipetting from tubes without manual transformation from tubes to plates. However, the TAT is longer than MGISP-960.
(2)MGISP-960 support 96 samples/run nucleic acid extraction with fixed throughput, however, its TAT is shorter than Smart 8.
The MGISP-Smart 8 is an open platform that enables users to customize hardware and scripts to meet their specific application needs. It supports third-party application kits, if customer wants to develop their own application, please contact local FAS for help.
Q12:Which certificates MGISP-Smart 8 has obtained at present?
So far, MGISP-Smart 8 has obtained many international certifications such as CE(RUO/IVD), CB, CMETus, EAC, UK, etc. It has access to many major markets such as EU, North America, UK and EAC, and the instrument quality is guaranteed .