Turbocharge Your Sequencing
Location:Home / Products

Turbocharge Your Sequencing

Powered by MGI's unique DNBSEQTM Technology, after fully upgrading biochemical, fluidics, and optical systems and launching High-throughput Sequencing Set V3.0, DNBSEQ-T7(1) can output 7 Tb of high-quality data within 24 hours and is widely applicable to Whole Genome Sequencing, Deep Exome Sequencing, Epigenome Sequencing, Transcriptome Sequencing, Tumour Panel, and other data-intensive projects.
Request Quote

Features

Features

High Speed: 22 hrs ~24 hrs for PE150 sequencing (V3.0*); 24 hrs ~30 hrs for PE150 sequencing (V2.0**)

High Flexibility: 4 Flow Cells, PE150, and PE100 at the same time

Ultra-high Throughput: 7 Tb (V3.0*) / 6 Tb (V2.0**) per day; high quality data around the clock


* DNBSEQ-T7RS High-throughput Sequencing Reagent Kit V3.0

** DNBSEQ-T7RS High-throughput Sequencing Reagent Kit V2.0

Performance Parameter

DNBSEQ-T7
Model DNBSEQ-T7
No. of Flow Cell per run 4
No. of lanes/Flow Cell 1
Max reads/Flow Cell* 5800 M** / 5000 M***
Read lengths

SE50

PE100

PE150

* The maximum number of effective reads are based on the sequencing of an internal standard library. Actual output may vary with sample types and library preparation methods.

**Max reads/Flow Cell based on using DNBSEQ-T7RS High-throughput Sequencing Reagent Kit V3.0

***Max reads/Flow Cell based on using DNBSEQ-T7RS High-throughput Sequencing Reagent Kit V2.0


DNBSEQ-T7
Product Name Reads Lengths Data Output Data QualityQ30* Run Time**
DNBSEQ-T7RS High-throughput Sequencing Set V3.0 PE100 1.16 Tb ~4.64 Tb >85% 16 hrs ~18 hrs
PE150 1.75 Tb ~7 Tb >85% 22 hrs ~24 hrs***
DNBSEQ-T7RS High-throughput Sequencing Set V2.0
PE100
1 Tb ~4 Tb
>85%
20 hrs ~22 hrs
PE150
1.5 Tb ~6 Tb
>85%
24 hrs ~30 hrs

* The percentage of base above Q30 is the average of an internal standard library over the entire run. The actual performance is affected by factors such as sample type, library quality, and insert fragment length.

** Run time includes Flow Cell loading, sequencing, base calling, etc. Cal. is a binary file format generated by MGI Sequencer basecall software.  

*** 24hrs sequencing time is based on using DNBSEQ-T7 High-throughput Sequencing Reagent Kit V3.0 and FPGA server.

System Parameter

  • Product Model*

    DNBSEQ-T7

    DNBSEQ-T7RS**

    *Only for model classification.

    **For research only. Not for use in diagnostic procedures.



  • Operating Environment Requirements

    Temperature: 19℃~25℃; fluctuation < 2°C per hour

    Relative Humidity: 30%RH ~ 80%RH; non-condensing

    Atmospheric Pressure: 80 kPa~106 kPa

    Waterproof Rating: IPX0

    Altitude: Below 2000 meters


  • Control Computer Configurations***

    CPU: Intel CORE I7-7700 4 Core×2 3.6 GHz

    Internal Storage: 16 Gb RAM
    HDD: 1 Tb
    SSD: 128 G
    Operating System: Windows 10

    ***Support the computer configurations and system updates

  • Power & Dimensions & Net Weight

    Power Type: 200 V~240 V, 50/60 Hz, 30 A

    Rated Power: 3000 VA

    Dimensions(L×W×H): 1656 mm×903 mm×1815 mm

    Net Weight: 765 Kg


  • Bandwidth for Network Connection

    300 Mb/s: For local storage network uploads

    4000 Mb/s: For Fastq computing uploads

    500 Mb/s: For Data analysis uploads


(1) Unless otherwise informed, this StandardMPS sequencing reagent is not available in Germany, UK, Sweden, and Switzerland.

  • Hotline (Global)
    +86-4000-688-114
  • Email
    MGI-service@mgi-tech.com
  • Special Recommendations
  • *For research use only.

    Not for use in diagnostic procedures (except as specifically noted).
Copyright © 2024 All Rights Reserved MGI Tech Co., Ltd. Guangdong ICP 16117185Privacy PolicyLegal NoticesCookie Policy
Our use of cookies
We would like to use necessary cookies to improve your browsing experience and the quality of our website. We would also like to set analytics cookies and advertisement cookies that help us make improvements by measuring how you use our website. Detailed information about the use of cookies on this website and how you can control your consent can be found in our Cookie Policy and Privacy Policy.
Accept only strictly necessary cookiesAccept all cookies