Small but Mighty: the Desktop Sequencer that Packs a Punch
Features
10 Gb~150 Gb per run
|
PE150 sequencing within only 40 hrs |
support several read lengths including SE50, PE50, SE100, PE100, and PE150
|
support sequencing applications for research, clinic, agriculture, etc
|
Performance Parameter
DNBSEQ-G50 |
||
---|---|---|
Flow Cell No. |
1 | |
Flow Cell Type |
FCS | FCL |
No. of Lanes per Flow Cell |
1 lane | |
Effective Reads |
100 M |
500 M |
Read Length |
SE100 PE100 PE150 |
SE50 PE50 SE100 PE100 PE150 |
* The maximum number of effective reads are based on the sequencing of an internal standard library. Actual output may vary with sample types and library preparation methods. |
DNBSEQ-G50 |
||||
---|---|---|---|---|
Read Lengths |
Reads |
Data Output |
Q30 * |
Run Time** |
SE50-FCL | 500 M | ~25 Gb | >85% | ~7 hrs |
PE50-FCL | 500 M | ~50 Gb |
>85% |
~15 hrs |
SE100-FCL | 500 M | ~50 Gb | >85% | ~13 hrs |
PE100-FCL | 500 M |
~100 Gb |
>85% |
~26 hrs |
PE150-FCL | 500 M | ~150 Gb | >80% | ~40 hrs |
SE100-FCS | 100 M | ~10 Gb | >80% | ~10 hrs |
PE100-FCS | 100 M | ~20 Gb | >85% | ~20 hrs |
PE150-FCS | 100 M | ~30 Gb | >80% | ~28 hrs |
* The percentage of base above Q30 is the average of an internal standard library over the entire run. The actual performance is affected by factors such as sample type, library quality, and insert fragment length. |
||||
** Run time is calculated over the entire run including sample loading, sequencing, base calling, and data processing. |
System Parameter
- Operating Environment Requirements*
Temperature: 19℃~25℃
Relative Humidity: 20% RH~80% RH, non-condensing
Atmospheric Pressure: 70 kPa~106 kPa
Waterproof Rating: IPX0
*For indoor use only.
- Control Computer Configurations*
CPU: 8 th Gen Core i7
Internal Storage: 32 Gb RAM
HDD: ≥4 Tb
Operating System: Windows 10 IOT (64-bit version)
*Support the computer’s configurations and system updates.
- Power & Dimensions & Net Weight
Type: 100 V~240 V, 50/60 Hz
Rated Power: 900 VA
Dimensions: length: 654 mm; width: 489 mm; height: 545 mm
Net Weight: 85 kg
- Product Model*
DNBSEQ-G50
DNBSEQ-G50 RS (For research only. Not for use in diagnostic procedures)
DNBSEQ-G50 CX (clinical purpose, overseas IVD)
Remark:
*means that “the product model names, DNBSEQ-G50 and MGISEQ-200, are used in the selected oversea countries and China, respectively.”
Related Products
- DNA Sequencing Library Preparation System MGISP-100RS
- High-throughut Automated Sample Preparation Systm MGISP-960RS
- MGIEasy Cell-free DNA Library Prep Set
- MGIEasy Single Cell Whole Genome Amplification Kit
- MGICare Detection Kit for Chromosome Copy Number Variation Test
- MGIEasy Rapid DNA Library Prep Kit customized for CDC
Related Cases
- ViewsDNBSEQ-G50 Helps With Study on the Transcriptome Structure of Novel CoronavirusOn March 15th, the RNA Research Center of IBS (Institute of Basic Science) of the Seoul National Un...
- ViewsHow to Decipher the Full-length Sequence of the New Coronavirus GenomeThe nucleic acid detection technology based on real-time fluorescent RT-PCR method has played an im...